ORBISPHERE K1100 LDO Sensor Kit, 0-2000 ppb, 28mm ORBISPHERE fitting, wall mount

ORBISPHERE K1100 LDO Sensor Kit, 0-2000 ppb, 28mm ORBISPHERE fitting, wall mount
Termékszám: K1100-KTO-W-MET
Egységár Elérhetőségünk
Elérhető

One calibration per year

One zero point calibration per year is all that is needed with the K1100 sensor. Designed for minimal drift, luminescent technology makes the K1100 sensor the most stable sensor with the longest calibration interval in the industry.

No membranes = two minutes of maintenance

With no membranes to replace and no electrolyte solution to replenish, the K1100 requires only two minutes of maintenance per year. Corrosive or hazardous chemicals are not required, making the annual task faster, easier and safer without reducing measurement precision.

Low cost retrofit

The complete system consists of a controller, a flow chamber, and the K1100 Luminescent Dissolved Oxygen Sensor. The sensor is compatible with Hach Orbisphere 28 mm flow chambers, eliminating the need for engineering changes. Installation is fast and easy and does not require special preparation.

A new level of confidence

The K1100 optical sensor is the first to use luminescent measurement technology to measure both ppb and ppm oxygen levels in power plants. Since 1978, Hach Orbisphere sensors have set the industry standard for oxygen measurement by delivering confidence to every water chemistry manager. The K1100 maintains this tradition and offers significant operating and cost benefits.

A doboz tartalma

Includes 410 wall mount controller, K1100 LDO sensor (low level), 3 m sensor cable, stainless steel flow-chamber with 6 mm fittings

Műszaki adatok

A doboz tartalma: Includes 410 wall mount controller, K1100 LDO sensor (low level), 3 m sensor cable, stainless steel flow-chamber with 6 mm fittings
Alkalmazás: Power
Analóg kimenetek: 3 Smart 0/4 to 20 mA (500 Ohms) programmable as linear or tri-linear, configurable to send diagnostics or alarm informations.
Áramellátási követelmények (feszültség): 100 - 230 V AC
Controller Options: 410 Mono-Channel Controller
Érzékelő: K1100 LDO
Felszerelési mód: 6 mm flow chamber
Garancia: 24 hónap
Hőmérséklet tartomány: Accurate from -5 - 50 °C
Resistant - temperature from -5 - 100 °C
Ismételhetőség: ± 0,4 ppb vagy 1 % amelyik nagyobb
Kalibrálás: Single point zero calibration with standard 99.999% nitrogen (quality 5.0) or equivalent oxygen free gas
Kijelző felbontás: 0.1 ppb
Kommunikációs képességek: 3 x 0/4-20 mA; RS485; Ethernet
Környezeti hőmérséklet: -5 - 50 °C
Legalacsonyabb detektálási határ: 0.6 ppb
Méréstartomány: 0 - 2000 ppb dissolved O₂ (DO)
Minta: nyomás: 1 - 20 bar absolute (14.5 - 290 psi)
Mintaáram: 50 - 300 mL/min
Minta hőmérséklet: -5 - 50°C
Paraméter: Oxygen
Precizitás (RDS): ± 0.8 ppb or 2 % whichever is greater
Relék: Measurement board: 3 measurement alarm relays (1A-30 VAC or 0.5A-50 VDC), configurable to send diagnostics information.
Main board: 1 system alarm relay (1A-30 VAC or 0.5A-50 VDC).
Reprodukálhatóság: ± 0,8 ppb vagy 2 % amelyik nagyobb
Tanúsítványok: 2004/108/EC - EN 61326-1
Tárolási feltételek: -5 °C - 100 °C
Tömeg szenzor: 0.6 kg
Válaszidő : (газ фаза)
Vezérlő: Wall

Egyéb információk a termékről