EZ7652 Total Nitrogen + Total Phosphorus Analyser

EZ7652 Total Nitrogen + Total Phosphorus Analyser
Termékszám: EZ7652.XXXXXXXX
Egységár Elérhetőségünk

A doboz tartalma

EZ7652 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)

Műszaki adatok

A doboz tartalma: EZ7652 Total Nitrogen & Total Phosphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 3 x empty 2.5L Reagent Containers with Fittings (HCl, NaOH and H₂SO₄ Solution) and 3 x empty 2.5L Glass Reagent Containers with Fittings (Colour, Reductor & Persulphate Solution)
Analóg kimenetek: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Anyag: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Automatikus tisztítás: Igen
Demineralised water: Consumption: 140 mL/analysis (hour)
Detektálási határ: TN: ≤ 0.5 mg/L
TP: ≤ 0.01 mg/L
Digitális kimenetek: Optional: Modbus (TCP/IP, RS485)
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Garancia: 24 hónap
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferenciák: TN: The main interferences are Br- and I-. When the amount of I- is 2.2 fold of the amount of TN, or the amount of Br- is 3.4 fold of the amount of TN, this will interfere on the test results. Dissolved organic matter, surfactants and Chromium (VI) interfere. Various inorganic substances not normally found in natural water, such as Chlorite [ClO₂-] and Chlorate [ClO₃-], may interfere.

TP: Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, and Vanadium (V), Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Kalibrálás: Automatic, 2-point; frequency freely programmable
Környezeti hőmérséklet: 10 - 30 °C, ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Lefolyócső: Atmospheric pressure, vented, min. Ø 32 mm
Mérés módszer: TN: UV photometric measurement at 220 nm after persulphate digestion in alkaline medium, based on APHA 4500-NO3 (B)
TP: Colorimetric measurement at 700 nm using ascorbic acid reduction and molybdate colour solution after persulphate digestion in acidic medium, based on APHA 4500-P
Méréstartomány: 0.5 - 20 mg/L TN; 0.010 - 2 mg/L TP
Méretek (MxSZxM): 690 mm x 465 mm x 330 mm
Minta: nyomás: By external overflow vessel
Mintaáram: 100 - 300 mL/min
Mintacellák: Standard measurement cycle time for both TN & TP: 60 minutes
Minta hőmérséklet: 10 - 30 °C
Minta minőség: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Mintautak száma: 1, 2, 4, or 8
Paraméter: Total Nitrogen (TN), Total Phosphorus (TP)
Pontosság: Better than 3% (TN) and 2% (TP) full scale range for standard test solutions
Reagent Requirements: Keep between 10 - 30 °C
Riasztás: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Súly: 25 kg
Tanúsítványok: CE compliant / ETL certified
Teljesítmény: 230 VAC, 50/60 Hz
120 VAC, 50/60 Hz
Max. power consumption: 440 VA
Validation: Automatic; frequency freely programmable
Védettségi osztály: Analyser cabinet: IP55 / Panel PC: IP65